|
Brief Instrument Overview |
|
We have two versions of the Microscope, a Room temperature and Low temperature Scanning Hall Probe Microscopes. Both microscopes can also be used as STM, AFM, and other Scanning Probes. In the standard configuration a sub-micron Hall probe microfabricated into a GaAs/GaAlAs heterostructure 2DEG is scanned over
the sample to measure the surface magnetic fields using conventional scanning tunneling microscopy-positioning techniques as shown in Fig. 1. A closer look at the microscope head is given at Fig. 2.
|
|
 Fig. 1 - Schematics of the Scanning Hall
Probe |
|
 Fig. 2.1 - Low Temperature SHPM head with sensors (STM tip, Quartz tuning fork, head)
 Fig. 2.2 - Details of the SHPM head |
|
It is expected that we should be able to reach the
highest magnetic measurement sensitivity currently available only with the
use of SQUIDs (~500µB) at helium temperatures. Our overall objective, is
to reach, in collaboration with Nanomagnetics, the sensitivity of several
Bohr magnetons (µB) and ultimately 1µB. This will be instrumental in
fabricating novel nanostructures, with potential application in
intelligent characterization of molecules and manipulation of single
charges and spins. The single spin detection capability will bring further
advances for nanoelectronics, nanomagnetics,
spintronics, and quantum information technology.
|
|
In order to approach the single spin sensitivity we are planning to:- Reduce the sample-probe separation to 50 nm (see estimates below).
- Decrease the minimum temperature from 4.2 to at least 1.2 K to reduce noise and improve the mobility of the 2D gas.
|
|
The minimum detectable magnetic field obtained with the SHPM, Bmin is given by the following formula: Bmin(T/√(Hz))=√(4kBTRs)/RHallIHall , where Rs, RHall and IHall are the series resistance, Hall resistance and the Hall current passing through the Hall probe, respectively. The Hall current cannot be increased indefinitely; there is a maximum current, IHall max , which can be passed through the probe without increasing the noise.
|
|
While the field sensitivity is naturally the highest at the lowest operating temperature (see the Table below), the scan size significantly decreases with temperature. Another trade off is the requirement that the surface should be conductive, as an STM feedback is presently used to scan immediately next to the sample surface. Magnetic field distribution of the sample is obtained simultaneously with the STM topography of the surface.
|
|
|